Type
|
Variable sampling volume,
total injection type automatic sample injector
|
|
Sample
Vial
|
250
§¡ PP
Optional: 2 §¢ glass, 96-well microplate, and 50 §¡ PP
|
|
Sampling
Microsyringe
|
Gas-Tight
Type 100 §¡
|
|
Sample
Injection Volume
|
100
§¡ Syringe: 1-80¥ìL, 0.1§¡ increment
0.1-0.8§¡, in 0.1§¡ steps
Maximum of 400§¡ injection possible with an optional sample loop
|
|
Injection
Precision
|
RSD
less than 1% at 1§¡ injection
|
|
Repeated
Injection
|
1-10
times for each vial
|
|
Max.
Sample Number
|
100 (100-sample vial holder, using 250§¡, 50§¡ PP vials)
Standard
200 (200-sample vial holder, using 250§¡, 50§¡ PP vials) Optional
50 (2 §¢
vial holder, using 2 §¢
glass vials) Optional
96 (96-well microplate holder, using 96-well microplate) Optional
|
|
Sample
Cooling
|
Electronic
Refrigeration, 4¡É-20¡É, Variable setting
|
|
Operational
Pressure
|
0-35
MPa
35 MPa (MAX)
|
|
Interruptible
Analysis
|
included
in Standard Model
|
|
Rinsing
Solution Port
|
included
in Standard Model
|
|
Error
Self-Diagnosis
|
included
in Standard Model
|
|
Input/Output
Signals
|
Integrator start, auto-zero, start, end, stop, error
signal, SYSCON
|
|
Power Supply
|
AC 100V ± 10%, 50/60 Hz, 350W
|
|
Dimensions
|
258(W) × 460(H) × 487(D) §®
|
|
Weight
|
31§¸
|